WarningRecall #Z-2001-2026Medical Device

Abbott Point Of Care Inc. Recalls i-STAT G3+ cartridge; List Number: 03P78-26;

Abbott Point Of Care Inc. · Class II — Moderate Risk

Hospital operating room medical equipment — representative photo for FDA medical device recall
A hospital operating room with medical equipment (representative file photo). Specific recalled device: see recall details below. Photo: FDA / Public Domain.

Product Description

i-STAT G3+ cartridge; List Number: 03P78-26;

Reason for Recall

Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.

FDA Classification

Class II — Moderate Risk

A situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, AUSTRIA, BANGLADESH, BELGIUM, FINLAND, FRANCE, GERMANY, IRELAND, ITALY, LUXEMBOURG, MAURITIUS, NETHERLANDS, NORWAY, POLAND, PORTUGAL, SPAIN, SWEDEN, SWITZERLAND, UNITED KINGDOM.

Remedy

Consumers should stop using this product. Contact the recalling firm for further instructions regarding this product.