Product Description

AccuVu Angiographic Catheter, Omni Flush, Non-Braided, With 21 R/O Markers; ACCU-VU OF 4F X 70CM 035 NB 6SH 21 R/O 1CM; Catalog No.: 13709704; Product/UPN No.: H787137097045 (Box), H787137097040 (Pouch); Box Quantity: 5 pouches;

Reason for Recall

The affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.

FDA Classification

Class II — Moderate Risk

A situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

Distribution

Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, China, Chile, Curacao, Czech Republic, Germany, Ireland, Israel, Italy, Lebanon, Netherlands, New Zealand, Panama, Slovenia, South Africa, Spain, Sweden, Thailand.

Remedy

Consumers should stop using this product. Contact the recalling firm for further instructions regarding this product.