Skip to content
411 PRESS
WarningRecall #Z-2049-2026Medical Device

Boston Scientific Corporation Recalls LUX-Dx II Plus

Boston Scientific Corporation · Class II — Moderate Risk

Hospital operating room medical equipment — representative photo for FDA medical device recall
A hospital operating room with medical equipment (representative file photo). Specific recalled device: see recall details below. Photo: FDA / Public Domain.

Product Description

LUX-Dx II Plus, Arrythmia detector and alarm, Model M312, with SERVER SW LATITUDE DRAGON EU (Europe), Model 6446, Version 7.5, or SERVER SW LATITUDE DRAGON AU (Australia/New Zealand), Model 6448, Version 7.5

Reason for Recall

For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.

FDA Classification

Class II — Moderate Risk

A situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

Distribution

Worldwide distribution - US Nationwide and the country of Belgium.

Remedy

Consumers should stop using this product. Contact the recalling firm for further instructions regarding this product.