Medline Industries Recalls Medline medical procedure kits
This is a Class II recall. This product may cause temporary health problems or pose a slight threat of a serious nature.
Medline Industries Recalls Medline medical procedure kits
Medline Industries, LP · Class II — Moderate Risk

Product Description
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. CATARACT PACK, Medline Kit Number/SKU DYNJ48725A; 2. OCULAR PACK, Medline Kit Number/SKU DYNJ64138B; 3. OCULOPLASTY PACK, Medline Kit Number/SKU DYNJ64572A; 4. OCULOPLASTY PACK, Medline Kit Number/SKU DYNJ64572B; 5. MICROSURGERY PACK, Medline Kit Number/SKU DYNJ67284A; 6. BASIC EYE PACK, Medline Kit Number/SKU DYNJ81411A; 7. SMR SINUS PACK, Medline Kit Number/SKU DYNJ84478; 8. FESS/SEPTO PACK, Medline Kit Number/SKU DYNJ85025; 9. PLASTIC EYE PACK, Medline Kit Number/SKU DYNJ89030; 10. NEURO, Medline Kit Number/SKU DYNJ902935K; 11. OCULAR, Medline Kit Number/SKU DYNJ909065; 12. OCULAR, Medline Kit Number/SKU DYNJ909065A.
Reason for Recall
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
FDA Classification
Class II — Moderate Risk
A situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
Distribution
Worldwide distribution.
Remedy
Consumers should stop using this product. Contact the recalling firm for further instructions regarding this product.
Sources: FDA openFDA Enforcement Database