Medtronic Recalls Medtronic Sphere-9 Catheter
This is a Class II recall. This product may cause temporary health problems or pose a slight threat of a serious nature.
Medtronic Recalls Medtronic Sphere-9 Catheter
Medtronic, Inc. · Class II — Moderate Risk

Product Description
Medtronic Sphere-9 Catheter, Model Number AFR-00001; catheter, percutaneous, cardiac ablation, for treatment of atrial flutter.
Reason for Recall
ventricular fibrillation (VF) during radiofrequency (RF) therapy with the Medtronic Sphere-9 catheter in patients who have a Biotronik Implantable Cardioverter Defibrillator (ICD) or Biotronik Cardiac Resynchronization Therapy-Defibrillator (CRT-D) implanted. The observation is caused by an interaction between the Sphere-9 catheter current flow and a safety feature built into Biotronik ICDs and CRT Ds. Medtronic recommends physicians use increased caution when using the Sphere-9 catheter for RF ablation therapy in patients implanted with a Biotronik ICD/CRT-D system, and to ensure external defibrillation systems are readily available.
FDA Classification
Class II — Moderate Risk
A situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
Distribution
Worldwide
Remedy
Consumers should stop using this product. Contact the recalling firm for further instructions regarding this product.
Sources: FDA openFDA Enforcement Database