Skip to content
411 PRESS
CriticalRecall #Z-1942-2026Medical Device

Ventec Life Systems Recalls VOCSN V+Pro package

Ventec Life Systems, Inc. · Class I — Most Serious

Hospital operating room medical equipment — representative photo for FDA medical device recall
A hospital operating room with medical equipment (representative file photo). Specific recalled device: see recall details below. Photo: FDA / Public Domain.

Product Description

VOCSN V+Pro package, REF: PRT-01198-000; VOCSN V+Pro unit, REF: PRT-01185-000

Reason for Recall

Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to reduced below intended levels of fraction of inspired oxygen delivery prior to and during ventilation and this oxygen-enriched environment may increase fire risk

FDA Classification

Class I — Most Serious

A situation in which there is a reasonable probability that the use of, or exposure to, a violative product will cause serious adverse health consequences or death.

Distribution

US: OH, GA, CA, MO

Remedy

Consumers should stop using this product. Contact the recalling firm for further instructions regarding this product.

411 Press Analysis

Ventec Life Systems, Inc. is recalling a medical device classified as Class I by the FDA — the highest risk level. The recall was initiated on March 25, 2026 affecting 42. Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to reduced below intended levels of fraction of inspired oxygen delivery prior to and during ventilation and this oxygen-enriched environment may increase fire risk. It was distributed in US: OH, GA, CA, MO. Consumers and healthcare professionals should exercise caution and discontinue use of the recalled product. Contact the recalling firm directly for return or disposal instructions.

Independent news on labor, safety, and accountability.