Philips North America Recalls Philips Ingenia Ambition S with MR Elastography (MRE)
This is a Class II recall. This product may cause temporary health problems or pose a slight threat of a serious nature.
Philips North America Recalls Philips Ingenia Ambition S with MR Elastography (MRE)
Philips North America · Class II — Moderate Risk

Product Description
Philips Ingenia Ambition S with MR Elastography (MRE). 1. Model Number (REF): 781359. 2. Model Number (REF): 782108. 3. Model Number (REF): 782139.
Reason for Recall
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
FDA Classification
Class II — Moderate Risk
A situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
Distribution
Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar, Netherlands, Pakistan, Palestine, Peru, Poland, Portugal, Russian Fed., Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Kingdom.
Remedy
Consumers should stop using this product. Contact the recalling firm for further instructions regarding this product.
411 Press Analysis
Philips North America has recalled a medical device that the FDA classified as Class II affecting 7 units. The recall was initiated on April 14, 2026. The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small. It was distributed internationally, including the United States. Class II recalls involve products that might cause a temporary health problem or pose only a slight threat of a serious nature. Consumers should discontinue use and contact the firm.